Method for in vitro assessment of straylight from intraocular lenses.

Biomed Opt Express 2015, vol. 6, issue 11

Ocular straylight has been measured by means of psychophysical methods over the years. This approach gives a functional parameter yielding a straight comparison with optically defined light scattering, and the point-spread-function. This is of particular importance when the effect of intraocular lenses (IOLs) on postoperative straylight is sought. An optical system for straylight measurements of IOLs was adapted to a commercial device (C-Quant, Oculus), which employs such psychophysical method. The proposed modifications were validated using light-scattering filters and some sample IOLs. The measurements were performed by 3 observers to prove that results are independent from straylight of the eye. Other applications will be discussed.